OUTPUT |
Analog Voltage ±10V
Channel 1: Integrated Particle Volume [LWC] Channel 2: Integrated Particle Surface Area [PSA] Channel 3: Effective Droplet Radius [Re] |
MEASUREMENTS |
Channel 1: 0.002-10 g/m³
Channel 2: 5-20,000 cm²/m³ |
DROPLET SIZE RANGE | 3-50 µm Diameter |
MEASUREMENT ACCURACY |
Channel 1: 5%
Channel 2: 5% Channel 3: 10% |
MEASUREMENT PRECISION |
Channel 1: 2%
Channel 2: 2% Channel 3: 5% |
MEASUREMENT RATE | 0 Hz to 200 Hz |
CALIBRATION | Light-Diffusing Disk |
SAMPLE VOLUME | 3.00 cm³ |
LIGHT SOURCE | Laser Diode, 780 nm |
DETECTORS | Silicon Photo-diode |
ENVIRONMENTAL RANGE |
Temperature -50ºC to 50ºC
Relative Humidity 0% to 100+% Icing Protected |
POWER REQUIREMENTS |
120 VAC, 60 Watt (220 VAC Optional)
Anti-Icing Heaters 120 VAC, 120 Watt |
PROBE SIZE |
Length 85 cm
Width 18 cm Height 52 cm |
PROBE WEIGHT | 10.4 Kg |
MOUNTING CONFIGURATION | On Vertical or Horizontal Pipe |